Biology Faculty Articles

Title

High-resolution melt analysis of the minisatellite D1S80: A potential forensic screening tool

Document Type

Article

Publication Date

2014

Publication Title

Electrophoresis

Keywords

D1S80, Forensic, High-Resolution melt, Minisatellite, SNP

ISSN

0173-0835

Volume

35

First Page

3020

Last Page

3027

Abstract

High-resolution melt (HRM) analysis of the VNTR region of the human D1S80 locus, a 16-bp repeat minisatellite from approximately 400 to over 700 bp in length, was investigated. A Qiagen Rotor-Gene Q using the Type-it PCR HRM kit was used to acquire HRM curves for 14 single, and 16 biallelic, dsDNA samples. The HRM analysis was applicable over a range of DNA concentrations; however the characteristics of the melt curve did depend on the forward and reverse primer ratio. Despite the large amplicon size and the similarities of the repeat sequences, it was possible to discriminate different genotypes. Heterozygotes were clearly different from the homozygous variants and even small differences in the repeat sequence could be differentiated. However, the melt analysis requires a high-resolution system with temperature resolution of 0.02°C or better in order to sort out differences in these large amplicons of near identical GC content (in this case 56%). HRM analysis of amplicons with large repeat sequences can be used as a means of comparing DNA fragments. Examination of multiple sequences can be used to differentiate DNA samples and demonstrate the potential of HRM analysis as a rapid and inexpensive prescreening technique in forensic applications.

DOI

10.1002/elps.201400143

This document is currently not available here.

Peer Reviewed

Find in your library

Share

COinS